23
SensoLite Nova
Code value can be adjusted digit by digit. After entering this menu function modify the value
of the first digit with the Up and Down buttons. By pressing the OK button, the first digit will
be set and the second digit will be flashing, indicating that now its value can be modified.
Use the Up and Down buttons to adjust the value of the second digit similarly to what you
did with the first one and then press OK when you are ready. Now the last digit should be
flashing. After you have set the value of the last digit as well, store the new code by pressing
OK. The meter will return to the test screen where the new code value will be displayed.
Compare this code value with the code value affixed to the strips' vial label. If they are the
same you are ready to perform blood glucose test, if not repeat the code setting procedure
again.
Note: If no buttons are pressed for 2 minutes, the meter automatically turns off and
the actual code value remains the same as before.