42
SensoLite Nova
Important!
If the instrument does not work after battery replacement: replace the batteries again,
press and hold the OK button down for about 15-20 seconds, then replace batteries.
If batteries are replaced in 60 seconds, date and time are not needed to be reset.
After 60 seconds, date and time should be reset.
Code number and test data are stored and they will not be lost due to battery replace-
ment. They will be available after battery replacement as well.
Electrostatic discharge may interfere with the meter. In case of unusual effects (turn
off, reset, errors) follow the instructions in the “Error messages and Troubleshooting”
chapter.
MAINTENANCE
If SensoLite Nova is used according to this “Instructions For Use”, only minimal mainte-
nance is necessary. However, to always have accurate test results, we recommend you to
perform the following operations:
If necessary, use a wet cloth for cleaning the whole surface of instrument.
After maintenance check functionality. For more details see chapter
SensoLite Nova system checking”.