SensoLite Nova
SENSOLITE NOVA METER EXPLAINED
SensoLite Nova indicates blood glucose concentration by checking the reaction between
chemical reagents and the blood drop on the test strip. The reaction triggers, the generation
of a current in the test strip's reagent zone and this current is conducted to the meter. The
current is in correlation with blood glucose concentration. Therefore the meter can determine
blood glucose concentration by measuring the extent of current flow.
The whole measuring process is controlled by a microprocessor inside the SensoLite Nova
meter. The microprocessor also controls the internal calibration of the device and if any
failure is detected in the operation, a relevant error message is displayed (see section „Error
messages and Troubleshooting”).
SensoLite Nova automatically stores measured data together with the current date and
time. The memory capacity is large enough to store the data of the last 500 tests. Stored
data can also be uploaded to a host computer via IR interface with the help of an additional
device named LiteLink (shipped separately upon request).